Thermo–Shaker TS-100 provides intensive mixing and temperature control of samples in microtest tubes or PCR plate. Functions of heating (up to +100°C) and mixing can be performed both simultaneously and independently, i.e. the unit implements three devices in one:
1. Shaker;
2. Dry–block Thermostat;
3. Thermo–Shaker.
TS-100 is used for DNA analysis sample preparation, for extraction of proteins, polysaccharides, lipids and other cellular components. Features of TS-100 meet the increased requirements of the user, including:
Heating source is a printed heating board (12 V). Mixing is provided by movement of orbital type.
The instrument is applicable in:
Temperature Calibration Function
With the help of the temperature calibration function the user can calibrate the unit approx. ±6% of the selected temperature to compensate differences in the thermal behaviour of tubes from different manufacturers.
Temperature setting range | +25°C … +100°C |
---|---|
Temperature control range | 5°C above ambient … +100°C |
Temperature setting resolution | 0.1°C |
Temperature stability | ±0.1°C |
Temperature accuracy at +37°C | ±0.5°C |
Average heating speed from +25°C to +100°C | 4°C/min |
Temperature uniformity over the block at +37°C | ±0.1°C |
Temperature uniformity over the block at +100°C | ±0.2°C |
Temperature calibration coefficient range | 0.936…1.063 (± 0.063) |
Speed control range | 250–1400 rpm |
Digital time setting | 1 min–96 hrs (1 min increment) |
Timer sound signal | + |
Orbit | 2 mm |
Display | LCD, 16 x 2 signs |
Microprocessor controlled temperature, mixing speed and operation time | + |
Maximum continuous operation time | 168 h |
Overall dimensions (W×D×H) | 220x240x90 mm |
Weight | 3.7 kg |
Input current/power consumption | 12 V, 3.5 A / 42 W |
External powe |